Comparison of Acute Thrombogenicity for Magnesium versus Stainless Steel Stents in a Porcine Arteriovenous Shunt Model.

MedStar author(s):
Citation: Eurointervention. 14(13):1420-1427, 2019 Jan 20.PMID: 29741484Institution: MedStar Heart & Vascular InstituteForm of publication: Journal ArticleMedline article type(s): Journal ArticleSubject headings: *Arteriovenous Fistula | *Thrombosis | Animals | Magnesium | Stainless Steel | Stents | SwineYear: 2019ISSN:
  • 1774-024X
Name of journal: EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of CardiologyAbstract: Aims The aetiology for reduced thrombogenicity of the Magmaris resorbable magnesium scaffold (RMS) when compared with the Absorb bioresorbable vascular scaffold remains unclear. We therefore investigated whether the Magmaris RMS has platelet-repelling properties by comparing its acute thrombogenicity with an equivalent stainless steel stent in an arteriovenous shunt model. Methods and Results An ex vivo porcine carotid jugular arteriovenous shunt was established and connected to Sylgard tubing containing the Magmaris RMS with sirolimus-eluting PLLA coating and an equivalent 316L stainless steel stent with sirolimus-eluting PLLA coating. Six shunts (2 shunt runs per pig) were run comparing the 2 scaffolds (n=9) in alternating order. Nested generalised linear mixed models were employed to compare variables between scaffold groups. Confocal fluorescent microscopy costaining CD61/CD42b demonstrated that the 316L equivalent stent had significantly greater platelet coverage of the total scaffold compared with Magmaris (5.8% vs. 2.8%, Rate ratio 2.21 [1.41, 3.47], p=0.012). Scanning electron microscopy demonstrated significantly greater thrombus deposition on the 316L equivalent stent as a percentage of the total scaffold compared with Magmaris (8.0% vs. 5.3%, p=0.009). Magmaris also had significantly less CD14 positive monocyte deposition and a trend toward less PM-1 positive neutrophil compared with the 316L equivalent stent. Conclusion Despite having identical scaffold characteristics regarding geometrical design, Magmaris had significantly less thrombogenicity and inflammatory cell deposition compared with the equivalent 316L stainless steel stent in a porcine arteriovenous shunt model. These data suggest resorbable magnesium scaffolds may have inherent properties that reduce adhesion of platelets and inflammatory cells.All authors: Acampado E, Adams L, Cheng Q, Finn AV, Gai J, Harder C, Hellinga DG, Joner M, Kolodgie FD, Lipinski MJ, Torguson R, Torii S, Virmani R, Waksman R, Zumstein PFiscal year: FY2019Fiscal year of original publication: FY2018Digital Object Identifier: Date added to catalog: 2018-06-19
Holdings
Item type Current library Collection Call number Status Date due Barcode
Journal Article MedStar Authors Catalog Article 29741484 Available 29741484

Aims The aetiology for reduced thrombogenicity of the Magmaris resorbable magnesium scaffold (RMS) when compared with the Absorb bioresorbable vascular scaffold remains unclear. We therefore investigated whether the Magmaris RMS has platelet-repelling properties by comparing its acute thrombogenicity with an equivalent stainless steel stent in an arteriovenous shunt model. Methods and Results An ex vivo porcine carotid jugular arteriovenous shunt was established and connected to Sylgard tubing containing the Magmaris RMS with sirolimus-eluting PLLA coating and an equivalent 316L stainless steel stent with sirolimus-eluting PLLA coating. Six shunts (2 shunt runs per pig) were run comparing the 2 scaffolds (n=9) in alternating order. Nested generalised linear mixed models were employed to compare variables between scaffold groups. Confocal fluorescent microscopy costaining CD61/CD42b demonstrated that the 316L equivalent stent had significantly greater platelet coverage of the total scaffold compared with Magmaris (5.8% vs. 2.8%, Rate ratio 2.21 [1.41, 3.47], p=0.012). Scanning electron microscopy demonstrated significantly greater thrombus deposition on the 316L equivalent stent as a percentage of the total scaffold compared with Magmaris (8.0% vs. 5.3%, p=0.009). Magmaris also had significantly less CD14 positive monocyte deposition and a trend toward less PM-1 positive neutrophil compared with the 316L equivalent stent. Conclusion Despite having identical scaffold characteristics regarding geometrical design, Magmaris had significantly less thrombogenicity and inflammatory cell deposition compared with the equivalent 316L stainless steel stent in a porcine arteriovenous shunt model. These data suggest resorbable magnesium scaffolds may have inherent properties that reduce adhesion of platelets and inflammatory cells.

English

Powered by Koha