Coronary artery bypass at a drug-eluting resorbable magnesium scaffold site in a porcine model.

MedStar author(s):
Citation: Cardiovascular Revascularization Medicine. 2022 Feb 28PMID: 35361569Institution: MedStar Health Research Institute | MedStar Heart & Vascular InstituteForm of publication: Journal ArticleMedline article type(s): Journal ArticleSubject headings: IN PROCESS -- NOT YET INDEXEDYear: 2022ISSN:
  • 1878-0938
Name of journal: Cardiovascular revascularization medicine : including molecular interventionsAbstract: BACKGROUND/PURPOSE: Coronary artery bypass graft (CABG) in the presence of metallic permanent stents may be challenging. We explored the feasibility of performing CABG in 1-year Magmaris resorbable scaffold implants in healthy Yucatan pigs.CONCLUSIONS: Our limited early feasibility study in a healthy animal demonstrates the practicality of CABG in a Magmaris scaffold after 1 year despite the persistence of visible degradation product, suggesting that the prior coronary device implant does not preclude future coronary surgery. Copyright © 2022. Published by Elsevier Inc.METHODS/MATERIALS: One animal received Magmaris 3.0 x 20.0-mm scaffolds in the left anterior descending (LAD), left circumflex (LCx), and right coronary arteries (RCA) under fluoroscopy. After 376 days, the animal underwent off-pump CABG at the scaffold site, including a single "end-to-side" anastomosis involving the right internal mammary artery (RIMA) to the LAD. The LAD was grafted in the territory of the previous scaffold as demonstrated by angiogram and the D1 and D2 landmarks.RESULTS: Digital radiography of the intact heart showed all degraded scaffolds widely expanded and uniformly positioned in the vessel wall in their expected locations. The proximal and distal radiopaque markers were clearly visible. The anastomosis was clearly identified by titanium branch clips, and the RIMA bypass was surrounded by connective tissue, which was further dissected before micro-computed tomography (CT) imaging. On the micro-CT images, calcification of the device was rare. The anastomosis site appeared intact, and the LAD at treatment site was widely open.All authors: Cohen JE, Finn A, Garcia-Garcia HM, Hellinga D, Kolodgie F, Virmani R, Waksman RFiscal year: FY2022Digital Object Identifier: Date added to catalog: 2022-05-11
Holdings
Item type Current library Collection Call number Status Date due Barcode
Journal Article MedStar Authors Catalog Article 35361569 Available 35361569

BACKGROUND/PURPOSE: Coronary artery bypass graft (CABG) in the presence of metallic permanent stents may be challenging. We explored the feasibility of performing CABG in 1-year Magmaris resorbable scaffold implants in healthy Yucatan pigs.

CONCLUSIONS: Our limited early feasibility study in a healthy animal demonstrates the practicality of CABG in a Magmaris scaffold after 1 year despite the persistence of visible degradation product, suggesting that the prior coronary device implant does not preclude future coronary surgery. Copyright © 2022. Published by Elsevier Inc.

METHODS/MATERIALS: One animal received Magmaris 3.0 x 20.0-mm scaffolds in the left anterior descending (LAD), left circumflex (LCx), and right coronary arteries (RCA) under fluoroscopy. After 376 days, the animal underwent off-pump CABG at the scaffold site, including a single "end-to-side" anastomosis involving the right internal mammary artery (RIMA) to the LAD. The LAD was grafted in the territory of the previous scaffold as demonstrated by angiogram and the D1 and D2 landmarks.

RESULTS: Digital radiography of the intact heart showed all degraded scaffolds widely expanded and uniformly positioned in the vessel wall in their expected locations. The proximal and distal radiopaque markers were clearly visible. The anastomosis was clearly identified by titanium branch clips, and the RIMA bypass was surrounded by connective tissue, which was further dissected before micro-computed tomography (CT) imaging. On the micro-CT images, calcification of the device was rare. The anastomosis site appeared intact, and the LAD at treatment site was widely open.

English