Hemodynamic Performance of Sutureless vs. Conventional Bioprostheses for Aortic Valve Replacement: The 1-Year Core-Lab Results of the Randomized PERSIST-AVR Trial.

MedStar author(s):
Citation: Frontiers in Cardiovascular Medicine. 9:844876, 2022.PMID: 35252408Institution: MedStar Heart & Vascular InstituteForm of publication: Journal ArticleMedline article type(s): Journal ArticleSubject headings: IN PROCESS -- NOT YET INDEXEDYear: 2022ISSN:
  • 2297-055X
Name of journal: Frontiers in cardiovascular medicineAbstract: Conclusion: At 1-year of follow-up of a PERSIST-AVR patient sub-group, the study showed comparable hemodynamic performance in the sutureless and the stented-valve groups, confirmed by independent echo core lab. Perceval sutureless prosthesis provides optimal sealing at the annulus with equivalent PVL and central regurgitation extent rates compared to sutured valves. Sutureless valves are therefore a reliable and essential technology within the modern therapeutic possibilities to treat aortic valve disease. Copyright © 2022 Fischlein, Caporali, Asch, Vogt, Pollari, Folliguet, Kappert, Meuris, Shrestha, Roselli, Bonaros, Fabre, Corbi, Troise, Andreas, Pinaud, Pfeiffer, Kueri, Tan, Voisine, Girdauskas, Rega, Garcia-Puente, De Kerchove, Lorusso and on behalf of the PERSIST-AVR Investigators.Methods: Perceval Sutureless Implant vs. Standard Aortic Valve Replacement (PERSIST-AVR) is a prospective, randomized, adaptive, open-label trial. Patients undergoing AVR, as an isolated or combined procedure, were randomized to receive a sutureless [sutureless aortic valve replacement (Su-AVR)] (n = 407) or a stented sutured [surgical AVR (SAVR)] (n = 412) bioprostheses. Site-reported echocardiographic examinations were collected at 1 year. In addition, a subgroup of the trial population (Su-AVR n = 71, SAVR = 82) had a complete echocardiographic examination independently assessed by a Core Lab (MedStar Health Research Institute, Washington D.C., USA) for the evaluation of the hemodynamic performance.Objective: Sutureless aortic valves are an effective option for aortic valve replacement (AVR) showing non-inferiority to standard stented aortic valves for major cardiovascular and cerebral events at 1-year. We report the 1-year hemodynamic performance of the sutureless prostheses compared with standard aortic valves, assessed by a dedicated echocardiographic core lab.Results: The site-reported hemodynamic data of stented valves and sutureless valves are stable and comparable during follow-up, showing stable reduction of mean and peak pressure gradients through one-year follow-up (mean: 12.1 +/- 6.2 vs. 11.5 +/- 4.6 mmHg; peak: 21.3 +/- 11.4 vs. 22.0 +/- 8.9 mmHg). These results at 1-year are confirmed in the subgroup by the core-lab assessed echocardiogram with an average mean and peak gradient of 12.8 +/- 5.7 and 21.5 +/- 9.1 mmHg for Su-AVR, and 13.4 +/- 7.7 and 23.0 +/- 13.0 mmHg for SAVR. The valve effective orifice area was 1.3 +/- 0.4 and 1.4 +/- 0.4 cm2 at 1-year for Su-AVR and SAVR. These improvements are observed across all valve sizes. At 1-year evaluation, 91.3% (n = 42) of patients in Su-AVR and 82.3% in SAVR (n = 51) groups were free from paravalvular leak (PVL). The rate of mild PVL was 4.3% (n = 2) in Su-AVR and 12.9% (n = 8) in the SAVR group. A similar trend is observed for central leak occurrence in both core-lab assessed echo groups.All authors: Andreas M, Asch FM, Bonaros N, Caporali E, Corbi P, De Kerchove L, Fabre O, Fischlein T, Folliguet T, Garcia-Puente J, Girdauskas E, Kappert U, Kueri S, Lorusso R, Meuris B, Pfeiffer S, Pinaud F, Pollari F, Rega F, Roselli EE, Shrestha ML, Tan E, Troise G, Vogt F, Voisine PFiscal year: FY2022Digital Object Identifier: Date added to catalog: 2022-05-11
Holdings
Item type Current library Collection Call number Status Date due Barcode
Journal Article MedStar Authors Catalog Article 35252408 Available 35252408

Conclusion: At 1-year of follow-up of a PERSIST-AVR patient sub-group, the study showed comparable hemodynamic performance in the sutureless and the stented-valve groups, confirmed by independent echo core lab. Perceval sutureless prosthesis provides optimal sealing at the annulus with equivalent PVL and central regurgitation extent rates compared to sutured valves. Sutureless valves are therefore a reliable and essential technology within the modern therapeutic possibilities to treat aortic valve disease. Copyright © 2022 Fischlein, Caporali, Asch, Vogt, Pollari, Folliguet, Kappert, Meuris, Shrestha, Roselli, Bonaros, Fabre, Corbi, Troise, Andreas, Pinaud, Pfeiffer, Kueri, Tan, Voisine, Girdauskas, Rega, Garcia-Puente, De Kerchove, Lorusso and on behalf of the PERSIST-AVR Investigators.

Methods: Perceval Sutureless Implant vs. Standard Aortic Valve Replacement (PERSIST-AVR) is a prospective, randomized, adaptive, open-label trial. Patients undergoing AVR, as an isolated or combined procedure, were randomized to receive a sutureless [sutureless aortic valve replacement (Su-AVR)] (n = 407) or a stented sutured [surgical AVR (SAVR)] (n = 412) bioprostheses. Site-reported echocardiographic examinations were collected at 1 year. In addition, a subgroup of the trial population (Su-AVR n = 71, SAVR = 82) had a complete echocardiographic examination independently assessed by a Core Lab (MedStar Health Research Institute, Washington D.C., USA) for the evaluation of the hemodynamic performance.

Objective: Sutureless aortic valves are an effective option for aortic valve replacement (AVR) showing non-inferiority to standard stented aortic valves for major cardiovascular and cerebral events at 1-year. We report the 1-year hemodynamic performance of the sutureless prostheses compared with standard aortic valves, assessed by a dedicated echocardiographic core lab.

Results: The site-reported hemodynamic data of stented valves and sutureless valves are stable and comparable during follow-up, showing stable reduction of mean and peak pressure gradients through one-year follow-up (mean: 12.1 +/- 6.2 vs. 11.5 +/- 4.6 mmHg; peak: 21.3 +/- 11.4 vs. 22.0 +/- 8.9 mmHg). These results at 1-year are confirmed in the subgroup by the core-lab assessed echocardiogram with an average mean and peak gradient of 12.8 +/- 5.7 and 21.5 +/- 9.1 mmHg for Su-AVR, and 13.4 +/- 7.7 and 23.0 +/- 13.0 mmHg for SAVR. The valve effective orifice area was 1.3 +/- 0.4 and 1.4 +/- 0.4 cm2 at 1-year for Su-AVR and SAVR. These improvements are observed across all valve sizes. At 1-year evaluation, 91.3% (n = 42) of patients in Su-AVR and 82.3% in SAVR (n = 51) groups were free from paravalvular leak (PVL). The rate of mild PVL was 4.3% (n = 2) in Su-AVR and 12.9% (n = 8) in the SAVR group. A similar trend is observed for central leak occurrence in both core-lab assessed echo groups.

English