TY - BOOK AU - Weissman, Neil J TI - Longitudinal Hemodynamics of Transcatheter and Surgical Aortic Valves in the PARTNER Trial PY - 2017/// KW - *Aortic Valve Stenosis/su [Surgery] KW - *Aortic Valve/su [Surgery] KW - *Bioprosthesis KW - *Heart Valve Prosthesis KW - *Hemodynamics KW - *Transcatheter Aortic Valve Replacement/mt [Methods] KW - Aged KW - Aged, 80 and over KW - Aortic Valve/dg [Diagnostic Imaging] KW - Blood Flow Velocity KW - Echocardiography KW - Female KW - Heart Valve Prosthesis Implantation/mt [Methods] KW - Humans KW - Longitudinal Studies KW - Male KW - Reoperation KW - Stroke Volume KW - MedStar Heart & Vascular Institute KW - Journal Article N2 - Conclusions and Relevance: This large, core laboratory-based study of transcatheter heart valves revealed excellent durability of the transcatheter heart valves and SAVR. Abnormal findings in individual patients, suggestive of valve thrombosis or structural deterioration, were rare in this protocol-driven database and require further investigation; Design, Setting, and Participants: In this study, we analyzed core laboratory-generated data from echocardiograms of all patients enrolled in the Placement of Aortic Transcatheter Valves (PARTNER) 1 Trial with successful TAVR or surgical AVR (SAVR) obtained preimplantation and at 7 days, 1 and 6 months, and 1, 2, 3, 4, and 5 years postimplantation. Patients from continued access observational studies were included for comparison; Importance: Use of transcatheter aortic valve replacement (TAVR) for severe aortic stenosis is growing rapidly. However, to our knowledge, the durability of these prostheses is incompletely defined; Interventions: Successful implantation after randomization to TAVR vs SAVR (PARTNER 1A; TAVR, n=321; SAVR, n=313), TAVR vs medical treatment (PARTNER 1B; TAVR, n=165), and continued access (TAVR, n=1996). Five-year echocardiogram data were available for 424 patients after TAVR and 49 after SAVR; Main Outcomes and Measures: Death or reintervention for aortic valve structural indications, measured using aortic valve mean gradient, effective orifice area, Doppler velocity index, and evidence of hemodynamic deterioration by reintervention, adverse hemodynamics, or transvalvular regurgitation; Objective: To determine the midterm hemodynamic performance of balloon-expandable transcatheter heart valves; Results: Of 2795 included patients, the mean (SD) age was 84.5 (7.1) years, and 1313 (47.0%) were female. Population hemodynamic trends derived from nonlinear mixed-effects models showed small early favorable changes in the first few months post-TAVR, with a decrease of -2.9 mm Hg in aortic valve mean gradient, an increase of 0.028 in Doppler velocity index, and an increase of 0.09 cm2 in effective orifice area. There was relative stability at a median follow-up of 3.1 (maximum, 5) years. Moderate/severe transvalvular regurgitation was noted in 89 patients (3.7%) after TAVR and increased over time. Patients with SAVR showed no significant changes. In TAVR, death/reintervention was associated with lower ejection fraction, stroke volume index, and aortic valve mean gradient up to 3 years, with no association with Doppler velocity index or valve area. Reintervention occurred in 20 patients (0.8%) after TAVR and in 1 (0.3%) after SAVR and became less frequent over time. Reintervention was caused by structural deterioration of transcatheter heart valves in only 5 patients. Severely abnormal hemodynamics on echocardiograms were also infrequent and not associated with excess death or reintervention for either TAVR or SAVR; Trial Registration: clinicaltrials.gov Identifier: NCT00530894 UR - https://dx.doi.org/10.1001/jamacardio.2017.3306 ER -