A Unique Case of Central Line Fracture Due to Laser Lead Extraction.
Citation: JACC. Case Reports. 4(24):101633, 2022 Dec 21.PMID: 36684034Institution: MedStar Washington Hospital Center | MedStar Washington Hospital CenterlDepartment: Cardiovascular Disease Fellowship | Clinical Cardiac Electrophysiology Fellowship | Electrophysiology | Medicine/Internal MedicineForm of publication: Journal ArticleMedline article type(s): Case ReportsYear: 2022ISSN:- 2666-0849
Item type | Current library | Collection | Call number | Status | Date due | Barcode |
---|---|---|---|---|---|---|
Journal Article | MedStar Authors Catalog | Article | 36684034 | Available | 36684034 |
Transvenous laser-assisted lead extraction is successful, with a low procedural complication rate for a wide range of indications. Here, we report a case of right internal jugular triple-lumen central venous catheter fracture and subsequent embolism to the right pulmonary artery during laser lead extraction that was successfully retrieved with a gooseneck snare. (Level of Difficulty: Advanced.). Copyright © 2022 The Authors.