Biomechanics of Biodegradable Subacromial Balloon Spacer for Irreparable Superior Rotator Cuff Tears: Study of a Cadaveric Model. (Record no. 4345)

MARC details
000 -LEADER
fixed length control field 04213nam a22005417a 4500
008 - FIXED-LENGTH DATA ELEMENTS--GENERAL INFORMATION
fixed length control field 190621s20192019 xxu||||| |||| 00| 0 eng d
022 ## - INTERNATIONAL STANDARD SERIAL NUMBER
International Standard Serial Number 0021-9355
040 ## - CATALOGING SOURCE
Original cataloging agency Ovid MEDLINE(R)
099 ## - LOCAL FREE-TEXT CALL NUMBER (OCLC)
PMID 31169580
245 ## - TITLE STATEMENT
Title Biomechanics of Biodegradable Subacromial Balloon Spacer for Irreparable Superior Rotator Cuff Tears: Study of a Cadaveric Model.
251 ## - Source
Source Journal of Bone & Joint Surgery - American Volume. 101(11):e49, 2019 Jun 05.
252 ## - Abbreviated Source
Abbreviated source J Bone Joint Surg Am. 101(11):e49, 2019 Jun 05.
253 ## - Journal Name
Journal name The Journal of bone and joint surgery. American volume
260 ## - PUBLICATION, DISTRIBUTION, ETC.
Year 2019
260 ## - PUBLICATION, DISTRIBUTION, ETC.
Manufacturer FY2019
265 ## - SOURCE FOR ACQUISITION/SUBSCRIPTION ADDRESS [OBSOLETE]
Publication status ppublish
266 ## - Date added to catalog
Date added to catalog 2019-06-21
520 ## - SUMMARY, ETC.
Abstract BACKGROUND: A subacromial balloon spacer is an option to treat irreparable rotator cuff tears. We hypothesized that the balloon would restore glenohumeral contact pressure, the acromion-humeral interval, and deltoid load to intact values after a simulated irreparable supraspinatus tear in a cadaveric model.
520 ## - SUMMARY, ETC.
Abstract CLINICAL RELEVANCE: This study supports further investigation of the balloon spacer in comparative clinical studies of surgical options for irreparable rotator cuff tears, with clinically relevant evaluation measures and observation periods.
520 ## - SUMMARY, ETC.
Abstract CONCLUSIONS: In a cadaveric model of an irreparable supraspinatus tear, a balloon spacer restored intact-state glenohumeral contact pressures at most abduction angles, lowered the humeral head, and increased deltoid load at postoperative time 0.
520 ## - SUMMARY, ETC.
Abstract METHODS: Fourteen cadaveric shoulders (mean age at the time of death, 67.9 years) were tested using a custom test frame. In this frame, glenohumeral contact pressure, the acromion-humeral interval, and deltoid load were measured using a digital sensor, a MicroScribe, and a spring scale, respectively. Test conditions included the intact shoulder, a small supraspinatus tear, supraspinatus repair, repair plus balloon, an irreparable supraspinatus tear (rotator cable-insufficient), and an irreparable tear plus balloon. Load was applied in a simulated neutral arm position (balanced) and active shoulder abduction (unbalanced).
520 ## - SUMMARY, ETC.
Abstract RESULTS: When the balloon was inflated over the irreparable supraspinatus tear in the balanced condition, glenohumeral contact pressure increased by 122% (p = 0.006) compared with that for the irreparable tear at 0degree of abduction and by 94% (p = 0.046) at 60degree. In the unbalanced condition, pressure decreased in the irreparable tear condition after the balloon was inflated, restoring pressure to close to that in the intact state. The balloon did not restore glenohumeral contact area to that in the intact shoulder in either the balanced or the unbalanced condition. The irreparable tear displaced the humeral head superiorly in the unbalanced condition, decreasing the acromion-humeral interval. The balloon moved the head inferiorly by a mean (and standard error of the mean) of 6.2 +/- 1.3 mm (p < 0.001) at 0degree of abduction, 4.4 +/- 1.3 mm (p < 0.001) at 30degree, and 3.0 +/- 0.8 mm (p < 0.001) at 60degree. The balloon increased the deltoid load after an irreparable tear by 8.2% (p = 0.022) at 0degree, 12.6% (p = 0.002) at 30degree, and 11.1% (p = 0.008) at 60degree.
546 ## - LANGUAGE NOTE
Language note English
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element *Arthroplasty/mt [Methods]
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element *Rotator Cuff Injuries/su [Surgery]
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Acromion/su [Surgery]
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Aged
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Aged, 80 and over
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Biomechanical Phenomena
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Cadaver
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Female
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Humans
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Male
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Middle Aged
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM
Topical term or geographic name entry element Rotator Cuff Injuries/pp [Physiopathology]
651 ## - SUBJECT ADDED ENTRY--GEOGRAPHIC NAME
Institution Curtis National Hand Center
651 ## - SUBJECT ADDED ENTRY--GEOGRAPHIC NAME
Institution MedStar Union Memorial Hospital
656 ## - INDEX TERM--OCCUPATION
Department Orthopaedic Surgery
657 ## - INDEX TERM--FUNCTION
Medline publication type Journal Article
700 ## - ADDED ENTRY--PERSONAL NAME
Local Authors Canham, R Bruce
700 ## - ADDED ENTRY--PERSONAL NAME
Local Authors Lobao, Mario H
700 ## - ADDED ENTRY--PERSONAL NAME
Local Authors Melvani, Roshan
700 ## - ADDED ENTRY--PERSONAL NAME
Local Authors Murthi, Anand M
700 ## - ADDED ENTRY--PERSONAL NAME
Local Authors Parks, Breny G
790 ## - Authors
All authors Abboud JA, Canham RB, Lobao MH, Melvani RT, Murthi AM, Parks BG
856 ## - ELECTRONIC LOCATION AND ACCESS
DOI <a href="https://dx.doi.org/10.2106/JBJS.18.00850">https://dx.doi.org/10.2106/JBJS.18.00850</a>
Public note https://dx.doi.org/10.2106/JBJS.18.00850
942 ## - ADDED ENTRY ELEMENTS (KOHA)
Koha item type Journal Article
Item type description Article
Holdings
Withdrawn status Lost status Damaged status Not for loan Collection Home library Current library Date acquired Total Checkouts Full call number Barcode Date last seen Price effective from Koha item type
          MedStar Authors Catalog MedStar Authors Catalog 06/21/2019   31169580 31169580 06/21/2019 06/21/2019 Journal Article

Powered by Koha