Fiber-reinforced fixation implant for proximal interphalangeal joint arthrodesis shows implant bio-integration at 1-year follow-up.

MedStar author(s):
Citation: Journal of Foot & Ankle Surgery. 2021 Jul 03PMID: 34247921Institution: MedStar Union Memorial HospitalForm of publication: Journal ArticleMedline article type(s): Journal ArticleSubject headings: IN PROCESS -- NOT YET INDEXEDYear: 2021Local holdings: Available online through MWHC library: 2003 - present, Available in print through MWHC library:1999-2007Name of journal: Foot and ankle surgery : official journal of the European Society of Foot and Ankle SurgeonsAbstract: BACKGROUND: A new bio-integrative fiber-reinforced implant (OSSIOfiber R Hammertoe Fixation Implant, OSSIO Ltd., Caesarea, Israel) was developed for proximal interphalangeal joint (PIPJ) correction-arthrodesis. The main purpose of this clinical study was to assess implant bio-integration at 1-year follow-up.CONCLUSIONS: This study demonstrates safe bio-integration of the newly developed fiber-reinforced implant at 1-year follow-up without negative side effects. Copyright (c) 2021. Published by Elsevier Ltd.METHODS: Twenty-four patients, previously treated for a Hammertoe deformity using the bio-integrative, fiber-reinforced implant, were enrolled in this follow-up study. One-year follow-up included clinical examination, patient reported outcomes, radiographs, Magnetic Resonance Imaging (MRI) and bio-integration scoring.RESULTS: Proximal interphalangeal joint (PIPJ) radiographic fusion rate was 92% (n = 22). MRI was analyzed for 24 (100%) patients. In 100% of patients (n = 24), the border between implant and surrounding tissue was scored as partially visible. There were no cyst formation or fluid accumulation findings. Mild bone edema was detected in 29% (n = 7) and is attributed to the chronic distribution of forces due to chronic abnormal gait and pasture. None of the edema findings were considered as adverse implant-related finding. The mean bio-integration score was 7.71 +/- 0.46.All authors: Cicchinelli LD, Miller S, Richter M, Stalc JFiscal year: FY2022Digital Object Identifier: Date added to catalog: 2021-07-26
Holdings
Item type Current library Collection Call number Status Date due Barcode
Journal Article MedStar Authors Catalog Article 34247921 Available 34247921

Available online through MWHC library: 2003 - present, Available in print through MWHC library:1999-2007

BACKGROUND: A new bio-integrative fiber-reinforced implant (OSSIOfiber R Hammertoe Fixation Implant, OSSIO Ltd., Caesarea, Israel) was developed for proximal interphalangeal joint (PIPJ) correction-arthrodesis. The main purpose of this clinical study was to assess implant bio-integration at 1-year follow-up.

CONCLUSIONS: This study demonstrates safe bio-integration of the newly developed fiber-reinforced implant at 1-year follow-up without negative side effects. Copyright (c) 2021. Published by Elsevier Ltd.

METHODS: Twenty-four patients, previously treated for a Hammertoe deformity using the bio-integrative, fiber-reinforced implant, were enrolled in this follow-up study. One-year follow-up included clinical examination, patient reported outcomes, radiographs, Magnetic Resonance Imaging (MRI) and bio-integration scoring.

RESULTS: Proximal interphalangeal joint (PIPJ) radiographic fusion rate was 92% (n = 22). MRI was analyzed for 24 (100%) patients. In 100% of patients (n = 24), the border between implant and surrounding tissue was scored as partially visible. There were no cyst formation or fluid accumulation findings. Mild bone edema was detected in 29% (n = 7) and is attributed to the chronic distribution of forces due to chronic abnormal gait and pasture. None of the edema findings were considered as adverse implant-related finding. The mean bio-integration score was 7.71 +/- 0.46.