Preclinical safety and electrical performance of novel atrial leadless pacemaker with dual-helix fixation.

MedStar author(s):
Citation: Heart Rhythm. 19(5):776-781, 2022 May.PMID: 35066177Institution: MedStar Heart & Vascular InstituteForm of publication: Journal ArticleMedline article type(s): Journal ArticleSubject headings: *Lipopolysaccharides | *Pacemaker, Artificial | Animals | Cardiac Pacing, Artificial | Equipment Design | Heart Atria | Humans | Prostheses and Implants | Sheep | Swine | Treatment OutcomeYear: 2022Local holdings: Available online through MWHC library: 2004 - presentName of journal: Heart rhythmAbstract: BACKGROUND: Complications associated with transvenous pacemakers, specifically those involving the lead or subcutaneous pocket, may be avoided with leadless pacemakers (LPs). The safety and efficacy of single-chamber right ventricular LPs have been demonstrated, but their right atrium (RA) use poses new design constraints.CONCLUSION: The novel atrial LP demonstrated successful implantation, with acceptable electrical performance, mechanical stability, and safety in a 12-week preclinical study. Copyright (c) 2022 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.METHODS: A new LP was designed with a dual-helix fixation mechanism specific to the RA anatomy. A 12-week preclinical ovine study was conducted to evaluate implant success, electrical performance, mechanical stability, and safety in vivo, with supporting benchtop measurements to quantify the mechanical forces needed for device retrieval and dislodgment.OBJECTIVES: The purpose of this study was to evaluate the implant success, electrical performance, and safety of a novel RA LP design in benchtop and preclinical studies.RESULTS: LPs were successfully implanted in all 10 ovine subjects with no complications. The pacing capture threshold improved significantly over time from implant to week 12 (1.1 +/- 0.7 V vs 0.4 +/- 0.2 V, P = .008). Sensing amplitudes and pacing impedances were stable from implant to week 12 (4.8 +/- 1.8 mV vs 6.0 +/- 1.9 mV, P = .160; and 393 +/- 77 OMEGA vs 398 +/- 65 OMEGA, P = .922, respectively). Gross pathology and microscopic histology revealed no adverse interactions and no evidence of device dislodgment or clinically significant myocardial perforation. Benchtop ex vivo porcine atrial tissue measurements revealed greater pull forces required to dislodge the LP vs transvenous active fixation lead (0.42 +/- 0.18 lbf vs 0.29 +/- 0.08 lbf, P = .020), and greater rotational forces required for deliberate extraction (0.28 +/- 0.04 lbf vs 0.14 +/- 0.07 lbf, P <.001).All authors: Badie N, Banker RS, Breeman K, Cantillon DJ, Doshi R, Eldadah Z, Knops RE, Ligon D, Nee P, Neuzil P, Rashtian M, Rippy MK, Victorine KOriginally published: Heart Rhythm. 2022 Jan 20Fiscal year: FY2022Digital Object Identifier: Date added to catalog: 2022-02-21
Holdings
Item type Current library Collection Call number Status Date due Barcode
Journal Article MedStar Authors Catalog Article 35066177 Available 35066177

Available online through MWHC library: 2004 - present

BACKGROUND: Complications associated with transvenous pacemakers, specifically those involving the lead or subcutaneous pocket, may be avoided with leadless pacemakers (LPs). The safety and efficacy of single-chamber right ventricular LPs have been demonstrated, but their right atrium (RA) use poses new design constraints.

CONCLUSION: The novel atrial LP demonstrated successful implantation, with acceptable electrical performance, mechanical stability, and safety in a 12-week preclinical study. Copyright (c) 2022 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

METHODS: A new LP was designed with a dual-helix fixation mechanism specific to the RA anatomy. A 12-week preclinical ovine study was conducted to evaluate implant success, electrical performance, mechanical stability, and safety in vivo, with supporting benchtop measurements to quantify the mechanical forces needed for device retrieval and dislodgment.

OBJECTIVES: The purpose of this study was to evaluate the implant success, electrical performance, and safety of a novel RA LP design in benchtop and preclinical studies.

RESULTS: LPs were successfully implanted in all 10 ovine subjects with no complications. The pacing capture threshold improved significantly over time from implant to week 12 (1.1 +/- 0.7 V vs 0.4 +/- 0.2 V, P = .008). Sensing amplitudes and pacing impedances were stable from implant to week 12 (4.8 +/- 1.8 mV vs 6.0 +/- 1.9 mV, P = .160; and 393 +/- 77 OMEGA vs 398 +/- 65 OMEGA, P = .922, respectively). Gross pathology and microscopic histology revealed no adverse interactions and no evidence of device dislodgment or clinically significant myocardial perforation. Benchtop ex vivo porcine atrial tissue measurements revealed greater pull forces required to dislodge the LP vs transvenous active fixation lead (0.42 +/- 0.18 lbf vs 0.29 +/- 0.08 lbf, P = .020), and greater rotational forces required for deliberate extraction (0.28 +/- 0.04 lbf vs 0.14 +/- 0.07 lbf, P <.001).

English

Powered by Koha