MARC details
000 -LEADER |
fixed length control field |
02616nam a22003737a 4500 |
008 - FIXED-LENGTH DATA ELEMENTS--GENERAL INFORMATION |
fixed length control field |
210628s20212021 xxu||||| |||| 00| 0 eng d |
022 ## - INTERNATIONAL STANDARD SERIAL NUMBER |
International Standard Serial Number |
1774-024X |
024 ## - OTHER STANDARD IDENTIFIER |
Standard number or code |
10.4244/EIJ-D-21-00012 [doi] |
024 ## - OTHER STANDARD IDENTIFIER |
Standard number or code |
EIJ-D-21-00012 [pii] |
040 ## - CATALOGING SOURCE |
Original cataloging agency |
Ovid MEDLINE(R) |
099 ## - LOCAL FREE-TEXT CALL NUMBER (OCLC) |
PMID |
34031024 |
245 ## - TITLE STATEMENT |
Title |
Incidence and Standardized Definitions of Mitral Valve Leaflet Adverse Events After Transcatheter Mitral Valve Repair: the EXPAND Study. |
251 ## - Source |
Source |
Eurointervention. 2021 May 25 |
252 ## - Abbreviated Source |
Abbreviated source |
EuroIntervention. 2021 May 25 |
253 ## - Journal Name |
Journal name |
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology |
260 ## - PUBLICATION, DISTRIBUTION, ETC. |
Year |
2021 |
260 ## - PUBLICATION, DISTRIBUTION, ETC. |
Manufacturer |
FY2021 |
260 ## - PUBLICATION, DISTRIBUTION, ETC. |
Publication date |
2021 May 25 |
265 ## - SOURCE FOR ACQUISITION/SUBSCRIPTION ADDRESS [OBSOLETE] |
Publication status |
aheadofprint |
266 ## - Date added to catalog |
Date added to catalog |
2021-06-28 |
520 ## - SUMMARY, ETC. |
Abstract |
AIMS: We aimed to report the findings of the expert panel and standardize definitions for LAE. |
520 ## - SUMMARY, ETC. |
Abstract |
BACKGROUND: An independent panel of experts reviewed all investigator-reported cases of mitral valve leaflet adverse events (LAE) after MitraClipTM NTR/XTR in the EXPAND Study. |
520 ## - SUMMARY, ETC. |
Abstract |
CONCLUSIONS: Mitral valve repair with MitraClipTM NTR/XTR is safe. The rate of LAE is lower than previously reported using older generation devices. The proposed definitions and findings will help differentiate leaflet injury from inadequate leaflet insertion and SLDA, and provide guidance to consistently diagnose LAE post MitraClipTM. |
520 ## - SUMMARY, ETC. |
Abstract |
METHODS: Standard definitions for different types of LAE were formulated and events adjudicated after detailed review by the expert panel. |
520 ## - SUMMARY, ETC. |
Abstract |
RESULTS: Enrolling centers reported LAE in 35 cases, 11 leaflet injuries (9 tear, 2 perforation) and 24 single leaflet device attachment (SLDA). The panel confirmed LAE in 20 cases (2.0% incidence), 18 patients had SLDA and 4 had leaflet injury (2 cases had both SLDA and injury). Leaflet injury occurred during device implant and resulted in surgical valve replacement or death. SLDA-alone events were identified during implant (n=2), pre-discharge (7) or at 30 days of follow-up (7) and were resolved ( 2+ residual MR) with additional clips in 75% of cases. |
546 ## - LANGUAGE NOTE |
Language note |
English |
650 ## - SUBJECT ADDED ENTRY--TOPICAL TERM |
Topical term or geographic name entry element |
IN PROCESS -- NOT YET INDEXED |
651 ## - SUBJECT ADDED ENTRY--GEOGRAPHIC NAME |
Institution |
MedStar Heart & Vascular Institute |
657 ## - INDEX TERM--FUNCTION |
Medline publication type |
Journal Article |
700 ## - ADDED ENTRY--PERSONAL NAME |
Local Authors |
Asch, Federico M |
790 ## - Authors |
All authors |
Asch FM, Grayburn PA, Kar S, Little SH, Mackensen GB, Maisano F, Rinaldi MJ, Sorajja P |
856 ## - ELECTRONIC LOCATION AND ACCESS |
DOI |
<a href="https://dx.doi.org/10.4244/EIJ-D-21-00012">https://dx.doi.org/10.4244/EIJ-D-21-00012</a> |
Public note |
https://dx.doi.org/10.4244/EIJ-D-21-00012 |
942 ## - ADDED ENTRY ELEMENTS (KOHA) |
Koha item type |
Journal Article |
Item type description |
Article |