Dual-chamber leadless pacing: Atrioventricular synchrony in preclinical models of normal or blocked atrioventricular conduction.

MedStar author(s):
Citation: Heart Rhythm. 2023 Apr 17PMID: 37075958Institution: MedStar Heart & Vascular InstituteForm of publication: Journal ArticleMedline article type(s): Journal ArticleSubject headings: IN PROCESS -- NOT YET INDEXED | Year: 2023Local holdings: Available online through MWHC library: 2004 - presentISSN:
  • 1547-5271
Name of journal: Heart rhythmAbstract: BACKGROUND: Dual-chamber leadless pacemakers (LPs) require robust communication between distinct right atrial (RA) and right ventricular (RV) LPs to achieve atrioventricular (AV) synchrony.CONCLUSION: Successful AV-synchronous, dual-chamber DDD(R) leadless pacing using a novel, continuous, wireless communication modality was demonstrated across variations in posture and rhythm in a preclinical model. Copyright � 2023. Published by Elsevier Inc.METHODS: RA and RV LPs were implanted and paired in 7 ovine subjects (4 with induced complete heart block). AV synchrony (% AV intervals <300 ms) and i2i communication success (% successful i2i transmissions between LPs) were evaluated acutely and chronically. During acute testing, 12-lead electrocardiographic and LP diagnostic data were collected from 5-minute recordings, in 4 postures and 2 rhythms (AP-VP and AS-VP, or AP-VS and AS-VS) per subject. Chronic i2i performance was evaluated through 23 weeks postimplant (final i2i evaluation period: week 16-23).OBJECTIVE: The purpose of this preclinical study was to evaluate a novel, continuous implant-to-implant (i2i TM) communication methodology for maintaining AV-synchronous, dual-chamber DDD(R) pacing by the 2 LPs.RESULTS: Acute AV synchrony and i2i communication success across multiple postures and rhythms were median [interquartile range] 100.0% [100.0%-100.0%] and 99.9% [99.9%-99.9%], respectively. AV synchrony and i2i success rates did not differ across postures (P = .59, P = .11) or rhythms (P = 1, P = .82). During the final i2i evaluation period, the overall i2i success was 98.9% [98.1%-99.0%].All authors: Badie N, Banker RS, Booth DF, Doshi RN, Exner DV, Hadadi CA, Ip JE, Knops RE, Ligon D, Nair D, Neuzil P, Rashtian M, Reddy VY, Yang WFiscal year: FY2023Digital Object Identifier: Date added to catalog: 2023-06-28
Holdings
Item type Current library Collection Call number Status Date due Barcode
Journal Article MedStar Authors Catalog Article 37075958 Available 37075958

Available online through MWHC library: 2004 - present

BACKGROUND: Dual-chamber leadless pacemakers (LPs) require robust communication between distinct right atrial (RA) and right ventricular (RV) LPs to achieve atrioventricular (AV) synchrony.

CONCLUSION: Successful AV-synchronous, dual-chamber DDD(R) leadless pacing using a novel, continuous, wireless communication modality was demonstrated across variations in posture and rhythm in a preclinical model. Copyright � 2023. Published by Elsevier Inc.

METHODS: RA and RV LPs were implanted and paired in 7 ovine subjects (4 with induced complete heart block). AV synchrony (% AV intervals <300 ms) and i2i communication success (% successful i2i transmissions between LPs) were evaluated acutely and chronically. During acute testing, 12-lead electrocardiographic and LP diagnostic data were collected from 5-minute recordings, in 4 postures and 2 rhythms (AP-VP and AS-VP, or AP-VS and AS-VS) per subject. Chronic i2i performance was evaluated through 23 weeks postimplant (final i2i evaluation period: week 16-23).

OBJECTIVE: The purpose of this preclinical study was to evaluate a novel, continuous implant-to-implant (i2i TM) communication methodology for maintaining AV-synchronous, dual-chamber DDD(R) pacing by the 2 LPs.

RESULTS: Acute AV synchrony and i2i communication success across multiple postures and rhythms were median [interquartile range] 100.0% [100.0%-100.0%] and 99.9% [99.9%-99.9%], respectively. AV synchrony and i2i success rates did not differ across postures (P = .59, P = .11) or rhythms (P = 1, P = .82). During the final i2i evaluation period, the overall i2i success was 98.9% [98.1%-99.0%].